Presence of biofeedback (biofeedback).
available
Presence of motivating 3D game environments
Наличие
Video recording of evaluation and training procedures
available
Appointment of the device
intended for: 1) analysis of biomechanical/neuro-muscular walking function using specialized equipment; 2) neuromuscular rehabilitation therapy for various anatomical areas based on patient interaction
Method of motion registration
biometric sensors
Automatic calculation of walking time parameters, walking phases, spatial parameters of walking, kinematic parameters of walking
available
Ability to pre-assess walking function prior to training.
available
Possibility to select any recorded walking parameter (temporal, spatial, EMG or kinematic) as B
available
Ability to change the complexity of training tasks during training.
available
Possibility of carrying out training in automatic mode (automatic selection of training difficulty)
available
Possibility to control other (not selected as parameters of the EBS) parameters of the patient's walking during training
available
Automatic generation of evaluation and training protocol
available
Saving completed assessments and drills, as well as their protocols in the database
available
Possibility to change the list of calculated parameters for biomechanical/neuromuscular function analysis
available
Ability to view previously completed walking function assessments/trainings and their protocols.
available
Ability to create custom templates for walking function assessment and training protocols
available
Built-in protocol template editor
available
Ability to edit evaluation and training protocols
available
Ability to include graphs, figures and tables in the protocols
available
Availability of a consolidated protocol to track the dynamics of rehabilitation
availability
Base of medical norms on basic parameters of walking (phases of walking, temporal parameters, spatial parameters, kinematics)
available
Monitoring the condition of equipment during evaluations and drills
available
Types of registration channels
accelerometer, gyroscopic, electromyographic
Sensor weight
no more than 40 (± 4) g
Dimensions of the sensor
not more than 60x40x17 (± 2) mm
Wireless communication with PC
availability
communication range
at least 10 m in line of sight
Maximum number of sensors simultaneously linked to one PC
at least 12 pcs
Sensor synchronization accuracy
discrepancy no more than 0.005 s
Quantization frequency
at least 200 Hz
Input range of measured accelerations in each of the three axes
not less than ±16g
Permissible error of acceleration measurement in each of the three axes
no more than ±(0.04 g + 5% of measured value)
Input range of angular velocities in each of the three axes
not less than ± 2000 deg/sec
Permissible error of angular velocity measurement in each of the three axes
no more than ±(0.5 deg/s + 2% of measured value)
Total number of differential channels per 1 sensor
at least 2
In-phase noise reduction (CMRR)
at least 100 dB
Lower bandwidth limit at level (-3 ± 0.5) dB in EMG measurement mode
no more than 5 Hz
Upper limit of the bandwidth at the level of (-3 ± 0.5) dB in the EMG measurement mode
at least 500 Hz
Natural noise level in EMG measurement mode
no more than 2 µV RMS
Input differential resistance
at least 100 mohm
EMG signal input range
not narrower than 0.07 µv – 20 mv
EMG signal measurement error in the input range from 0.07 µV to 20 mV and at the bandwidth from 5 to 500 Hz
no more than ±(7 µV + 5% of the measured value)
Suppression of mains interference by a switchable tap selector filter
at least 40 dB
Function for determining the sub-electrode impedance
Available