It is used to resuscitate a patient after childbirth or to maintain positive pressure during artificial ventilation or weakened breathing.
Provides formation of oxygen-air mixture with adjustable oxygen percentage, gas mixture flow and pressure with protection against exceeding the limit value. Provides indication of output pressure and flow.
Fully pneumatic (no power supply required).
Flow rate with standard rotameter with operating range up to 10 litres/min.
Inlet working pressure range 200-600 kPa
Range of oxygen percentage in the mixture 21 - 100 %
Oxygen-air mixture mixing accuracy ± 3 %
Alarm if the pressure difference between air and oxygen supplied to the product exceeds > 300 kPa
Built-in filters with check valve in the inlet channels of the 60m product
Output pressure range0 - 90 cm H2O
Pressure limit range, max90 cm H2O
Product dimensions (WxHxH) mm200x...