Input signal
60 mV, 75 mV, 100 mV, 150 mV, 200 mV, 250 mV, 500 mV, 1000 mV, 2000 mV; 1 V, 2 V, 5 V, 10 V, 2…10 V, 20 V, 50 V, 100 V, 150 V, 200 V, 250 V, 300 V, 500 V, 750 V; 1 mA, 2 mA, 5 mA, 10 mA, 20 mA, 4…20 mA, 50 mA, 100 mA, 200 mA, 500 mA, 1000 mA, 2000 mA; 1 A,
Basic error
по току и напряжению: ±0,1 %, ±0,2 %; по аналоговому выходу: ±0,5 %
Overall dimensions, mm
72x72x103 (with rear cover), 72x72x75.6 (without cover) mm
LED indication
1 block of seven-segment indicators for displaying measured signal values by measuring channel; single LED indicators showing interface operation, state of discrete outputs, highlighting the unit of measurement; height z
Range of readings
from -19999 to +19999; from -9999 to +9999 (interface-less version)
Registering the maximum measured value
yes
Galvanic isolation of input and output circuits, power supply circuits
yes
RS485 communication interface
Quantity: 0, 1, 2; Modbus RTU protocols; Interface baud rate: 4800, 9600, 19200, 38400, 57600, 115200 bps.
Analog outputs
quantity: 0, 1, 2; ranges: 0...5 mA, 4...20 mA, 0...20 mA, 0...2.5...5 mA, 4...12...20 mA, 0...10...20 mA
Discrete outputs
Quantity: 0, 1, 2 Constant voltage 350 V, 200 mA or AC voltage 250 V, 200 mA
Power supply voltage
5 Vn – (5 + 4/-0.5) V DC; 12 Vn – (12 + 6/-3) V DC; 24 Vn – (24 + 12/-6) V DC; 220 VU - from 85 to 305 V AC at 50 Hz or from 100 to 430 V DC; 230 V - from 85 to 305 V AC at
Power consumption from the power supply circuit, not more
2.5 VA (for devices 5V, 12V, 24V); 4.0 VA (for devices without RS485 220V, 230V); 5.0 (for devices 220V, 230V)
reprogramming
via 'Configurator' program (RS485 interface) or using front panel buttons (if available)
Operating temperature range
From -40 to +70 °C (relative humidity 98% at +35 °C)
Degree of protection
IP54
Inspection interval
10 years
versions
general industrial; for use at nuclear power plants (safety class - 4 according to NP-001-2015)
Warranty period of operation
60 months
Average service life, at least
minimum 20 years