performance
at least 120 patient tests/h for glucose only. At least 80 patient tests/h for simultaneous glucose, lactate, and hemoglobin determination.
Simultaneous loading of samples
up to 6 samples in the carriage
Sensor type
combined type sensor. Simultaneous measurement of glucose and lactate by one sensor. Ceramic plate with enzymes protected by casing. Optionally, only glucose measurement is possible.
Positions in rotor
up to 4 cells, one for: calibrator tube, urgent sample, norm control and pathology control.
Availability of an inbuilt barcode scanner for tubes
available
Sample types
whole venous and capillary blood, serum, plasma, cerebrospinal fluid
Sample volume
No more than 20 µl Pediatric sample no more than 10 µl
Measuring range
not less than 0.5 to not more than 50.0 mmol/l for glucose; not less than 0.5 to not more than 40.0 mmol/l for lactate not less than 1.2 to not more than 20 mmol/l for hemoglobin
Measuring accuracy
< 1.5% for glucose < 1.5% for lactate < 1.5% for hemoglobin
Types of sensors used
1. sensor for glucose determination; 2. sensor for glucose and lactate determination; 3. sensor for hemoglobin determination
Calibration modes
1. Automatic calibration 2. User-requested calibration
Integrated monitoring program for all automatic functions
available
Automatic detection and indication of system errors
available
Automatic recognition of the presence of a test tube
available
Photometric corrector, wavelength 530 nm
availability. Designed to obtain plasma glucose concentration value when measuring whole blood samples.
Pump type
high-precision piston pump in glass housing, silent, no parts requiring regular replacement
Thermal sensor
If the temperature goes beyond the working range (from +15 to +35°C), a corresponding message appears on the display and printer
Automatic fluid level control
early warning for quick replacement of system solution and drain containers
Results output
to display and printer
menu
fully localized with color images
Energy saving sleep mode
presence
Possibility of complete shutdown of the device while keeping the current sensor functional
available
Number of stored measurement results
100 pcs
Number of protocols stored in the memory
100 pcs
Weight of photometric unit
0.05 kg