Type of digital protocol
HART v.7 (4...20 mA)
version
general industrial, explosion-proof 'intrinsically safe electrical circuit i'
Accuracy class
Up to 0.05 C (depending on type of NSH)
Possibility of simultaneous operation with two TCs using 2-wire scheme or with two TPs using built-in CHS
yes
Maximum measured temperature
up to +1820 °C
Minimum measured temperature
from minus 210 °C
TC connection diagrams
2-, 3-, 4-wire
Compliance with GOST R IEC 61508-2-2012 "Functional safety of electrical, electronic, programmable electronic systems".
Safety integrity level 2 (SIL 2)
Mounting of the measuring transducer (MT)
in temperature converter housing or on metal bar 'AD-01' for mounting on DIN rail
Explosion protection
0Ex ia IIA T6 Ga X, 0Ex ia IIB T6 Ga X, 0Ex ia IIC T6 Ga X, 0Ex ia IIA T5 Ga X, 0Ex ia IIB T5 Ga X, 0Ex ia IIC T5 Ga X, 0Ex ia IIA T4 Ga X, 0Ex ia IIB T4 Ga X, 0Ex ia IIC T4 Ga X, 0Ex ia IIA T3 Ga X, 0Ex ia IIB T3 Ga X, 0Ex ia IIC T3 Ga X, [Ex ia Ga] IIA X
Type (NCC) of primary converter
KVD (ISH), 50M, 100M, 50P, 100P, Pt100, Pt500, Pt1000, 100N, 1000N, TJK(J), TMK (T), THKn (E), THK(L), THA(K), TPP(S), TPP (S), TPR(B), THN(N), TVR (A-1), -100 mV...+100 mV, -1000 mV...+1000 mV, 400 Ohm, 4000 Ohm, 10 kOhm
Vibration resistance according to GOST R 52931-2008
group V3
Climatic version
UHL3.1 according to GOST 15150-69 (from -60 to +70 °C)
Degree of protection against dust and moisture
IP20
Power supply voltage
=16...42 B
Overall dimensions
diameter 57 mm, height 21 mm
Electromagnetic compatibility (EMC)
IIIA
Inspection interval
up to 4 years
Warranty period
24 months (from commissioning) or 36 months (from shipment)
Programmable error current
"Low" (3.7...3.8 mA) or "High" (20...23 mA)