Length
220 mm
Width
260 mm
Height
210 mm
View
Monitor of vital functions of the patient
Power supply voltage
198...242 V
Prohibitions of audible alarms
availability
Overall dimensions,mm
Not more than 260x210x220 mm
Accuracy class
not applicable
Electromagnetic compatibility
GOST R EC 60601-1-2
Defibrillator discharge protection
GOST R MEC 60601-1
Type of working parts (degree of protection of the patient against electric shock when electrodes are applied to the patient)
CF
Electrical safety protection class
Protection class II
Compliance with SAFETY standards
GOST R MEC 60601-1, GOST R MEC 60601-2-49
Computer software that allows the collection, processing, storage and visualization of monitored data,
availability
Integrated electronic help system
availability
Displaying the current date and time on the screen
availability
Quickly set audible alarm limits using the touch screen without navigating through submenus
availability
Virtual AD Quick Start button on the touch screen
availability
BP measurement quick start button on the front panel of the device
availability
Configuration of monitoring parameters by the user (enabling and disabling measurement channels at the user's choice)
availability
Monitoring mode on the screen at least 6 ECG leads simultaneously
availability
Number of displayed curves in standard mode
at least 6
Number of displayed curves in the "large digits" mode
Not less than 3
Big Digit Mode
availability
Saves all settings including operating modes and trends in the monitor memory in case of power failure
availability
Blocking the alarm sound for 4 minutes and resuming it after the specified time has elapsed
availability
Visual indication and audible signaling in case of ECG electrode breakage or poor quality electrode placement
availability
Visual indication and audible signaling in case of cardiac arrest (asystole) for a time exceeding (5 ± 0.3) s
availability
Visual indication and audible signaling in case of measured values of physiological parameters exceeding the set limits
availability
Setting the upper and lower alarm limits for all monitored parameters
availability
Absolute error limits of CO2 measurement under normal conditions at pressure 100-150 mmHg.
not regulated
Absolute error limits of CO2 measurement in normal conditions at pressure 70-100 mm.Hg.
-8...8
Absolute error limits of CO2 measurement under normal conditions at pressure 40-70 mm.Hg.
-5...5
Limits of permissible absolute error of CO2 measurement in normal conditions at pressure 0-40 mm.Hg.
-2...2
Range of measurement of CO2 in exhaled breath
0...150
Inhalation CO2 measurement range
0...50
Permissible absolute error of temperature measurement in the range from 32 to 42 C
-0.2...0.2 °C
Temperature measuring range (normalized)
32...42 °C
Temperature display range
25...50 °C
Number of temperature measurement channels
2 pc
Temperature measurement method
Thermistor
TRAINING mode for automatic adjustment of the breathing channel to recognize breathing cycles
Naliye
Selecting the breath sampling leads: RF or RL
availability
Range of values of variable component of impedance at which pneumogram registration in the range of respiratory frequency is pro
1...10 ohm
Range of basic impedance values at which pneumogram registration is provided in the respiratory rate range from 3 to 9
500...1000 ohm
Changes the vertical size of the PG within the limits (display scale)
1-10
Permissible absolute error of PD measurement
± 2 min-1
Breathing rate (HR) measurement range by ECG channel (impedance method) (normalised error)
3-90
Display range of respiratory rate (RR) by ECG channel (impedance method)
3-150
Displaying the time of the last BP measurement on the screen
availability
Display on the screen the remaining time until BP measurement (at the set measurement interval)
availability
Display the measurement mode (adult, pediatric, or neonatal) on the monitor screen
availability
Continuous BP measurement mode (user selectable)
availability
Automatic detection of the inflation level in the cuff (referenced to the previous measurement)
availability
Manual setting of neonatal pump level
60...120
Manual inflation level setting in adult mode
100...180
Automatic cuff inflation level setting from 100 to 280 mmHg.
availability
Starting the BP meter at any time at the user's initiative
availability
Automatic start-up of the BP meter with adjustable time interval between starts of 3, 5, 10, 15, 30 and 60 minutes
availability
Relative permissible error of cuff pressure measurement in the range from 150 to 280 mmHg.
-2...2 %
Absolute permissible error of cuff pressure measurement in the range from 20 to 150 mmHg.
-3...3
Cuff pressure measurement range
0...280
Measurement of SpO2 and HR at low perfusion levels
availability
Audible pulse accompaniment via photoplethysmogram channel (user-activated)
availability
Display of photoplethysmogram with automatic switching of sensitivity.
availability
Permissible absolute error of PD measurement
Within ± 2 beats/min.
Heart rate (HR) measurement range, beats/minute
30-240
Permissible absolute error of SpO2 measurement in the range of SpO2 values from 50 to 69 %
-3...3 %
Permissible absolute error of SpO2 measurement in the range of SpO2 values from 70 to 100%
-2...2 %
Saturation (SpO2) measurement range
50...100 %
ECG Isolation Stabilization Filter
availability
Using 3 electrodes (R,L,N) without connecting other electrodes to the ECG cable to capture only one ECG lead.
availability
Removable electrode holders in the ECG cable
availability
Selecting the HR lead: I, II or III and connecting other electrodes to the ECG cable to capture only one ECG lead.
availability
ECG channel protection against defibrillator pulses
availability
ECG QRS complex audio (user-activated)
availability
ECG waveform sweep speed
12,5, 25, 50 mm/second
Time for measuring and displaying new HR values
Not more than 20 seconds
Permissible absolute error of HR measurement (at R-beat amplitude from 0.2 to 4 mV)
Within ± 1 beats/min.
Heart rate (HR) measurement range (normalized for error), beats/min.
30-240
Heart rate (HR) display range, beats/minute
15-300
Direct current in the patient's circuit flowing through any electrode
Not more than 0,1 µA
ECG signal vertical display scale (sensitivity equivalent)
4 levels: 5, 10, 20 and 40 mm/mV
Input voltage range of ECG channel signals, mV
0.1-5
Output of ECG, PG, FPG curves and measured values of HR, BP, SpO2, HR, HR, HR, , T to a personal computer or central station.
availability
Entering, saving patient data (name, ID number, gender, age, bed number, onset and duration of observation), and storing patient
availability
Display of values (trends) of measured parameters in tabular form with intervals of 1, 5, 10, 15, 30 or 60 min (optional)
availability
Displaying values (trends) of measured parameters in graphical form
availability
Memorization of values (trends) of measured parameters for at least 24 hours with an interval of 1 minute
availability
Setting of alarm limits and alarms for all measured ??????????? causing an alarm condition
availability
Capnogram (CAP) display on the display screen.
availability
Displaying a photoplethysmogram (PPG) on the display screen
availability
Displaying a pneumogram (PG) on the display screen
availability
Display of electrocardiogram (ECG) in six leads (I, II, III, aVR, aVL, aVF) on the display screen
availability
Measure and display patient body temperature values (T1 and T2) on two channels
availability
Measurement and display of inhalation and exhalation CO2 values
availability
Impedance method measurement and display of respiratory rate (RR) values
availability
Measurement and display of saturation (SpO2) and heart rate (HR) values by photoplethysmography
availability
Measurement and display of non-invasive systolic (CAD), diastolic (DBP) and mean (MAP) blood pressure values
availability
Ability to select the lead for ECG HR measurements
availability
Measuring and displaying heart rate (HR) values from the ECG
availability
Type of backup power supply
Built-in batteries
Backup power supply operation in case of external power supply failure
availability
Weight (with built-in batteries)
3 kg
Weight (without built-in batteries)
2.5 kg
Power consumption of the monitor from AC mains power supply
40
Monitor mount type
Screw fixing holes in electronic unit housing
Mounting the monitor on a shelf, rack, in the car interior
availability
Mechanism for changing the tilt angle of the monitor when mounted on a flat surface
Extendable legs
Auxiliary controls (audible alarm blocking, start-stop BP measurement, call up trend view, call up menu of basic functions, call
Buttons on the front panel
Basic control (all monitor functions)
Touch screen
User Interface
Fully graphic, window, interactive
Screen resolution
At least 800x600 pixels
Screen Color Palette
24-bit TrueColour
Screen backlighting
dimmable LED lamp
Screen type
Touch - TFT LCD
Screen size
8
Carrying handle
availability
Housing material
Shatterproof plastic
Electronic unit design
Version 2
Design solution
Electronic unit
Age contingent of patients
Adults, children, newborns
Monitoring standard
Pulse oximetry, non-invasive blood pressure, electrocardiography, respiratory rate, body temperature, capnography