For sterilisation with saturated steam under overpressure of medical products made of metals, glass, rubbers, plastics, dressing materials, textile products, the impact of steam on which does not cause changes in their functional properties.
Set temperature regimes
132°C - 20 min,
120°C - 45 min.
110°C - 180 min.
It is equipped with a safety valve set to operate at overpressure, MPa 0,30±0,02.
Residual humidity for solid loading does not exceed, % 0,2.
Residual humidity for porous loading does not exceed, % 1.
Time of continuous operation per day, hours, not more than 16.
Average service life, years 10.
Number of sterilisation boxes of KSCF-18 type, simultaneously loaded into the chamber - 3 pcs. (supplied by separate order).
The warranty period of steriliser operation is 12 months from the moment of completion of commissioning works, but not more than 24 months from the date of manufacture.